Troubleshooting; Mattress Heating; Height Adjustment (Optional) - Dräger Medical Babytherm 8000 WB Instructions For Use Manual

Warming bed
Table of Contents

Advertisement

Troubleshooting

Mattress heating

Message
o
Red LED ±1
C flashing,
measured value display flashing,
intermittent alarm on.
o
Red LED 40
C flashing,
measured value display flashing,
continuous alarm on.
N
Red LED
lit,
continuous alarm on.
Red LED Inop lit.
The measured value display shows Err
alternately with the measured value.
The error code is flashing in the setting
display.
Continuous alarm sounds.
The three middle segments of the
measured value display are flashing.
The measured value display shows Acc
alternately with the measured value.
Intermittent alarm on.
Measured value display shows:
Three dashes at bottom
Three dashes at top

Height adjustment (optional)

Message
Height adjustment inoperative.
Cause
Measured value deviates from set value
o
by more than ±1
C.
Mattress temperature above 40
Power failure
Malfunction
Temperature sensors defective.
Power failure alarm battery is
insufficiently charged.
Mattress temperature is under 5
Mattress temperature is above 45
Cause
Motor overheat switch has tripped.
Remedy
When using in combination with radiant
heater: Reduce the output level of the
radiant heater.
o
C
When using in combination with radiant
heater: Reduce the output level of the
radiant heater.
Check that unit is plugged into mains.
Check that mains is switched on.
Call DrägerService.
Call DrägerService.
Call DrägerService.
Operate unit for 30 minutes,
press G key to mute alarm
for 15 minutes.
o
C
Wait until mattress temperature
o
is above 5
C
o
C
Wait until mattress temperature
o
is under 45
C
Remedy
Allow to cool for about 15 minutes.
Troubleshooting
31