CPI Canada Inc.
3C.6.0 APR EDITOR (Cont)
Definition of the APR EDITOR function.
FUNCTION
(MEMBRANE
CONSOLE)
APR EDITOR
NOTE:
APR text may be altered by using a computer running GenWare
with GenWare
This function does not apply to GenWare®.
Use these steps to set the APR EDITOR. Refer to the definition in the previous table.
Step
1.
From the GENERATOR SETUP menu, select APR EDITOR.
2.
Toggle the button to select ENABLED or DISABLED.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
DESCRIPTION
Enables / disables the ability of the operator to make and then save changes to APR techniques.
Allows the operator to change the default APR technique(s), and then save the changes to
ENABLED:
memory. The APR will subsequently default to the changed technique.
Allows temporary editing of APR technique(s), but does not allow the changes to be saved to
DISABLED:
memory. The APR will always default to the original technique when the generator is switched
OFF and then ON again.
The generator stores the last APR EDITOR setting before being switched off. If the APR editor was
previously ENABLED, APR changes may subsequently be made and then saved in normal operating
mode without the need to manually set the APR editor to ENABLED. To disable APR technique
changes, the APR editor must be set to DISABLED.
®
in the form of an MS Word document.
Ch # 901476-04
®
. Further documentation regarding this function is included
Action (membrane console)
Rev. C