CPI CMP 200 Series Service Manual page 304

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
8.2.0
SPARE PARTS LIST (Cont)
2.
The AEC board in your generator was selected to be compatible with specific AEC devices. To
maintain full compatibility, the original part number must be ordered as a replacement. Refer to
chapter 9, section 9.2.0, for the part number of the original AEC board that was shipped in the
generator for which this manual was prepared.
3.
This applies to the dual speed starter option only, spares should be stocked accordingly. The part
number shown for the dual speed starter board is for the board only, without the phase-shift
capacitors. For the complete dual speed starter subassembly, several part numbers (which are tube-
stator dependent) are used in CMP 200® DR 400 / 480 VAC generators. To determine which dual
speed starter subassembly is in your generator, note the DUAL SPEED STARTER ASSY part number
on a label near the top left side of the dual speed starter chassis. This will be the part number that
must be ordered for spares usage. For reference, this will be part number 901297-XX or 901298-XX
where XX is a two digit number designating the exact configuration.
4.
This generator control board is jumper configurable. Please refer to the following tables to configure
the generator control board for your x-ray generator.
WARNING:
FUNCTION
Firmware
Upgrading
kV Overvoltage
Trip
Cooling Fan
Maximum kV
DAP
HT Tank
Feedback Comp.
32 / 40 kW
400 / 480 VAC
125 kV
JW4
Pins 1-2
JW5
Pins 1-2
5.
These assemblies apply to generators equipped with a membrane console only.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
DO NOT ATTEMPT TO RECONFIGURE 125 kV MAXIMUM GENERATORS FOR 150 kV
OPERATION. SERIOUS GENERATOR DAMAGE WILL OCCUR.
JUMPER
JW2
JW4 and JW5
JW6
JW7
JW12
JW13
JW14
JW15
50 kW
400 / 480 VAC
150 kV
Pins 2-3
Pins 1-2
TABLE 8-5
PINS 1-2: Normal
PINS 2-3: Upgrade
Set as per Table 8-5b below.
PINS: 1-2: 125 kV
PINS: 2-3: 150 kV
Not used.
PINS 1-2: Fan
PINS 2-3: No Fan
PINS 1-2: 125 kV (For 125 kV tank)
PINS 2-3: 150 kV (For 150 kV tank)
Set as per DAP Installation in Chapter 3C.
PINS 1-2: 125 kV
PINS 2-3: 150 kV
TABLE 8-5b
32 / 40 kW
32 / 40 kW
208 / 230 VAC
400 / 480 VAC
125 kV
150 kV
Pins 2-3
Pins 1-2
Pins 2-3
Pins 1-2
Ch # 901476-10
CONFIGURATION
50 kW
208 / 230 VAC
150 kV
Pins 2-3
Pins 2-3
Rev. C
8
Spares
65 kW
400 / 480 VAC
150 kV
Pins 2-3
Pins 2-3
Page 8-7