CPI CMP 200 Series Service Manual page 106

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
3C.7.1 Tube Selection / Tube Setup (Cont)
Step
Action (membrane console)
Select FIL BOOST. Use the + or –
12.
buttons to set the filament rapid boost
duration.
Select FIL PREHEAT. Use the + or –
13.
buttons to set the filament preheat time.
Press >>.
14.
Select MAX SF MA. Use the + or –
15.
buttons to set the maximum small focus
mA.
16.
Select ANODE HU WARNING. Use the +
17.
or – buttons to set desired anode HU
warning %.
Select ANODE HU LIMIT. Use the + or –
18.
buttons to set desired anode HU limit %.
Press Return or alternatively << to exit to
19.
the Tube 1 Selection screen.
Press EXIT as required to return to the
20.
GEN CONFIGURATION menu.
Typically, the boost time should be between 200 and 250 msec, and the preheat time should be in the range of 700 - 800 ms.
If in doubt, monitor the filament feedback and be sure that there is no change in the signal level 5 ms. after the start of an exposure, and that
the mA starts at the selected level.
Standby current must be below the emission point.
If the maximum filament current is increased, be careful not to exceed the tube manufacturer's specifications.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Action (PC GenWare®)
Set the filament rapid boost duration via
the Filament Boost Time (ms) dialog
box.
Set the filament preheat time via the
Filament Preheat Time (ms) dialog box.
Select the Tube Settings tab.
Set the maximum small focus mA via the
Max. SF Current (mA) dialog box.
Select the Anode Heating tab.
Set the desired anode HU warning % via
the Anode Warning (%) dialog box.
Set the desired anode HU limit % via the
Anode Limit (%) dialog box.
Ch # 901476-04
Rev. C
Interfacing and Programming
Action (Touchscreen GenWare®)
Set the filament rapid boost duration via
the Filament Boost Time (ms) dialog
box.
Set the filament preheat time via the
Filament Preheat Time (ms) dialog box.
Select the Tube Settings tab.
Set the maximum small focus mA via the
Max. SF Current (mA) dialog box.
Set the desired anode HU warning % via
the Anode Warning (%) dialog box.
Set the desired anode HU limit % via the
Anode Limit (%) dialog box.
Page 3C-37