CPI CMP 200 Series Service Manual page 265

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
5.2.3
Error Messages (Cont)
ERROR
MESSAGE
CODE
(membrane console)
E103
CAL-MAN. TERM.
E104
CAL-NO MA
E105
CAL-MIN MA ERR.
E150
TABLE COMM ERR
E152
E154
E240
OUT OF LABELS
E241
LABEL JAMMED
E242
PRINTER ERROR
E243
PRINTER COMM ERR
E244
PLATEN OPEN
E245
PRINTER OFF-LINE
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
PROBLEM
(touch screen console)
Calibration error –
Operator released exposure button
manually terminated
during auto calibration.
Calibration error – no mA
No mA feedback detected during
auto calibration.
Calibration error –
Minimum generator mA was
minimum mA not
exceeded at start of calibration.
calibrated
Table communication
A communication error has occurred
error
with the table.
Table error
The table has sent a table error
message to the generator.
Table emergency stop
The table emergency stop has been
activated.
The printer is out of labels.
The labels are jammed.
The printer self-diagnostics have
reported a printer problem.
The printer has reported a
communication error.
The paper platen is not positioned
properly.
The printer is off-line.
Ch # 901476-07
Rev. B
5
Troubleshooting
ACTION
Retry auto Calibration.
Call product support.
Reduce filament standby
current on primary and/or
secondary filament.
1. Ensure that the table is
powered on.
2. Check the serial
communication cable,
connectors, etc.
Refer to the table
manufacturers documentation.
Reset the table emergency
stop.
Load more labels.
Clear the paper jam.
Refer to the printer manual.
Try printing again.
Check the platen position.
Put the printer on-line.
Page 5-12