3C.13.10 Touchscreen Console Firmware Upgrade / Restore (Cont)
Step
Once the communication media is selected, the file path, version and media information will appear in the Server Information
6.
window of the CPI Touch Screen Updater.
From the Upgrade Media Selection window select Upgrade via a USB to Computer Connection.
7.
The CPI Touch Screen Updater will read the available firmware packets (see figure 3C-42).
8.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Figure 3C-41: CPI Touch Screen Updater - Select Communication Media screen
Ch # 901476-04
Action
Rev. C
Interfacing and Programming
Page 3C-117
3C