Reproducibility, Linearity, & HVL Testing
5.0
H.V.L. EVALUATION (Cont)
ADDED ALUMINUM FILTER
0
0
0
0 (Average of above three readings)
2
(total 2 mm)
1
(total 3 mm)
3
(total 6 mm)
By interpolating the thickness of Al at a relative transmission value of 0.5, it can be seen that the HVL is
approximately 3.3.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 12
DOSE (mGy)
2.17
2.13
2.16
2.15
1.41
1.14
0.61
Table 12: HVL dose values (example)
Figure 3: HVL plot (example)
Rev. C
X-Ray Generator Service Manual Supplement
RELATIVE TRANSMISSION
1.00
.66
.53
.28