CPI CMP 200 Series Service Manual page 260

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
5.2.3
Error Messages (Cont)
ERROR
MESSAGE
CODE
(membrane console)
E017
NOT CALIBRATED
E018
PREP TIMEOUT
E019
ANODE HEAT LIMIT
E020
THERMAL INT #1
E022
DOOR INTERLOCK
E028
PREP SW CLOSED
E029
X-RAY SW CLOSED
E032
CONSOLE COMM
ERR
E033
GEN BATTERY LOW
E034
+12VDC ERROR
E035
-12VDC ERROR
E038
CAL DATA ERROR
E039
AEC DATA ERROR
E041
REC DATA ERROR
E042
TUBE DATA ERR.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
PROBLEM
(touchscreen console)
Uncalibrated Exposure
Selected mA not calibrated for
Parameter
selected kV.
Preparation Timeout
Generator has been in prep state too
long.
Anode Heat Limit
Selected parameters will cause X-
ray tube to exceed its programmed
anode heat limit.
Thermal Switch Interlock
X-ray tube # 1 too hot and its thermal
#1 Error
switch has opened.
Door Interlock Error
Door is open.
Prep Input Active During
Prep input active during power on
Initialization Phase
initialization phase.
X-ray Input Active During
X-ray input active during power on
Initialization Phase
initialization phase.
Console Communication
Generator has detected error in
Error
communication to console.
Warning Lithium Battery
Generator detects lithium battery
Voltage Low
voltage is low.
+12 V DC Error
+12VDC rail is out of tolerance.
-12 V DC Error
-12VDC rail is out of tolerance.
Calibration Data Corrupt
Generator detects corrupt calibration
Error
data.
AEC data corrupt
Generator detects corrupt AEC data.
Receptor data corrupt
Generator detects corrupt receptor
data.
Tube data corrupt
Generator detects corrupt tube data.
Ch # 901476-07
Rev. B
5
Troubleshooting
ACTION
Recalibrate X-ray tube.
Reduce length of time in prep
state.
Reduce parameters or wait for
tube to cool.
Wait for X-ray tube # 1 to cool.
Close door.
Check prep switch and input
for short circuit.
Check X-ray switch and input
for short circuit.
1. Check console cable for
damage and proper
connection.
2. Turn power off and then on
to reset generator.
Replace lithium battery.
Check +12VDC rail.
Check -12VDC rail.
Re-calibrate X-ray tube(s).
Reprogram AEC data or set
factory defaults.
Reprogram receptor data or
set factory defaults.
Reprogram tube data or set
factory defaults.
Page 5-7