3C
Interfacing and Programming
3C.5.3 Statistics (Cont)
Use these steps to view the generator statistics.
Step
Action (membrane console)
1.
From the UTILITY menu, select
STATISTICS.
2.
The STATISTICS menu displays
exposure data, and allows the exposure
counter to be reset as described below:
The TUBE 1 EXP counter displays
the tube 1 exposure count made
since this counter was last reset.
TOTAL EXP displays the total rad
exposure count. This is not resettable
*
Select RESET TUBE 1 EXP to reset
the tube 1 exposure counter.
*
The subject exposure counter is reset when the factory defaults are reset. Therefore, the "total exposure" count should be
recorded before resetting the factory defaults.
3.
Press EXIT to return to the UTILITY
menu.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-18
Action (PC GenWare®)
Select Generator Statistics from the
Utility menu, or use the statistics button
on the GenWare® toolbar.
The Generator Statistics window
displays exposure data, and allows the
exposure counter to be reset as
described below:
The Tube 1 Rad Exposures counter
displays the tube 1 exposure count
made since this counter was last
reset.
Total Number of Rad Exposures
displays the total rad exposure count.
*
This is not resettable
.
Select Reset Tube 1 to reset the tube
1 exposure counter.
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Action (Touchscreen GenWare®)
Use the statistics button
GenWare® toolbar to access the
Generator Statistics window.
The Generator Statistics window
displays exposure data, and allows the
exposure counter to be reset as
described below:
The Tube 1 Rad Exposures counter
display the tube 1 exposure count
made since this counter was last
reset.
Total Number of Radiographic
Exposures displays the total rad
exposure count. This is not resettable
*
.
Select Reset Tube 1 to reset the tube
1 exposure counter.