3C.7.5 AEC Setup (Cont)
Use these steps to perform the initial AEC setup. Refer to the definitions in the previous table.
Step
Action (membrane console)
From the GEN CONFIGURATION menu,
1.
select AEC SETUP.
2.
From AEC SETUP menu 1, select the
3.
AEC channel to be programmed.
Pressing the CHANNEL button will scroll
through the available AEC channels.
4.
Select LEFT FIELD. Toggle the button to
select YES or NO.
5.
Repeat the above for the CENTER and
RIGHT fields.
6.
Select CHAMBER TYPE. Use the + or –
buttons to select the desired AEC
chamber type.
7.
8.
Select FILM SCREEN 1. Toggle the
button to select YES or NO.
9.
Repeat the above for FILM SCREEN 2
and FILM SCREEN 3.
10.
Press >>.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Action (PC GenWare®)
From the GenWare® GENERATOR
UTILITIES application, select AEC Setup
and Calibration from the Setup menu, or
use the AEC Setup & Calibration button
on the GenWare® toolbar.
Select the AEC Setup tab.
Under Channel Number, select the AEC
channel that is to be programmed.
Click the left field on the graphic under
Fields to enable / disable that field. Black
indicates that the selected field is
disabled; green indicates that the
selected field is enabled.
Repeat the above for the center and right
fields.
Under Chamber Type, select the desired
AEC chamber type.
Select ION if using solid state AEC board assembly 737992.
Select the Film Screen options tab.
Under Film Screen Active, select film
screen 1. An X in the check box indicates
that film screen 1 is enabled.
Repeat the above for film screen 2 and
film screen 3.
Select the AEC setup tab.
Ch # 901476-04
Rev. C
Interfacing and Programming
Action (Touchscreen GenWare®)
Press the
button on the
GenWare® toolbar to access the AEC
Setup and Calibration utility.
Select the AEC Setup tab.
Under Channel Number, select the AEC
channel that is to be programmed.
Click the left field on the graphic under
Fields to enable / disable that field. Black
indicates that the selected field is
disabled; green indicates that the
selected field is enabled.
Repeat the above for the center and right
fields.
Under Chamber Type, select the desired
AEC chamber type.
Select the Film Screen options tab.
Under Film Screen Active, select film
screen 1. An X in the check box indicates
that film screen 1 is enabled.
Repeat the above for film screen 2 and
film screen 3.
Select the AEC setup tab.
Page 3C-65