CPI CMP 200 Series Service Manual page 257

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
5.2.2
Limit Messages (Cont)
MESSAGE
(membrane console)
GEN DUTY WARNING
GEN. JOULE LIMIT
GEN. KV LIMIT
GEN. KW LIMIT
GEN. MA LIMIT
GEN. MAS LIMIT
GEN. MS LIMIT
INVALID PARAM.
TUBE KV LIMIT
TUBE KW LIMIT
TUBE MA LIMIT
TUBE MAS LIMIT
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
(touchscreen console)
Generator duty cycle
The X-ray generator has reached its duty cycle
limit
warning limit.
Generator Joule Limit
The requested exposure will exceed the
generators kilo joule heat limit.
Generator kV limit
Requested kV not allowed as generator kV limit
has been reached.
Generator kW limit
Requested parameter not allowed as generator
kW limit has been reached.
Generator mA limit
Requested mA not allowed as generator mA
limit has been reached.
Generator mAs limit
Requested mAs not allowed as generator mAs
li130
mit has been reached.
Generator ms limit
Requested ms not allowed as generator ms limit
has been reached.
Invalid communication
Generator detected invalid parameter within
parameter.
received message, message ignored.
Parameter limit
The selected parameter has exceeded its limits.
Tube kV limit
Requested kV not allowed as tube kV limit has
been reached.
Tube kW limit
Requested parameter not allowed as tube kW
limit has been reached.
Tube mA limit
Requested mA not allowed as tube mA limit has
been reached.
Tube mAs limit
Requested mAs not allowed as tube mAs limit
has been reached.
Ch # 901476-07
PROBLEM
Rev. B
5
Troubleshooting
ACTION
Re-evaluate technique factors.
Allow generator to cool if
possible. If exposures are
continued, serious generator
damage may result due to
overheating.
Allow the generator to cool
sufficiently
to
allow
the
requested exposure.
None.
None.
None.
None.
None.
Select valid parameter.
None.
None.
None.
None.
None.
Page 5-4