3C
Interfacing and Programming
3C.7.1 Tube Selection Tube Setup (Cont)
Step
Action (membrane console)
Select MAX SF KW LS. Use the + or –
4.
buttons to set the low speed, small focus
kW limit.
Select MAX LF KW LS. Use the + or –
5.
buttons to set the low speed, large focus
kW limit.
Select MAX SF KW HS. Use the + or –
6.
buttons to set the high speed, small focus
kW limit.
Select MAX LF KW HS. Use the + or –
buttons to set the high speed, large focus
kW limit.
7.
Select MAX KV. Use the + or – buttons to
8.
set the maximum allowable kV.
Press >>.
9.
Select SF STANDBY. Use the + or –
buttons to set the small focus filament
standby current.
Select LF STANDBY. Use the + or –
buttons to set the large focus filament
standby current.
Select SF MAX. Use the + or – buttons to
10.
set the small focus maximum filament
current.
Select LF MAX. Use the + or – buttons to
11.
set the large focus maximum filament
current.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-36
Action (PC GenWare®)
Set the low speed, small focus kW limit
via the Max SF Power (kW) dialog box,
under Low Speed.
Set the low speed, large focus kW limit
via the Max LF Power (kW) dialog box,
under Low Speed.
Set the high speed, small focus kW limit
via the Max SF Power (kW) dialog box,
under High Speed.
Set the high speed, large focus kW limit
via the Max LF Power (kW) dialog box,
under High Speed.
Select the Tube Settings tab.
Set the maximum allowable kV via the
Max. Tube Voltage (kV) dialog box.
Select the Filament Parameters tab.
Set the small focus filament standby
current via the Standby Current (A)
dialog box, under Small Focus.
Set the large focus filament standby
current via the Standby Current (A)
dialog box, under Large Focus.
Set the small focus maximum filament
current via the Max Current (A) dialog
box, under Small Focus.
Set the large focus maximum filament
current via the Max Current (A) dialog
box, under Large Focus.
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Action (Touchscreen GenWare®)
Set the low speed, small focus kW limit
via the Max SF Power (kW) dialog box,
under Low Speed.
Set the low speed, large focus kW limit
via the Max LF Power (kW) dialog box,
under Low Speed.
Set the high speed, small focus kW limit
via the Max SF Power (kW) dialog box,
under High Speed.
Set the high speed, large focus kW limit
via the Max LF Power (kW) dialog box,
under High Speed.
Select the Tube Settings tab.
Set the maximum allowable kV via the
Max Tube Voltage (kV) dialog box.
Select the Filament Parameters tab.
Set the small focus filament standby
current via the Standby Current (A)
dialog box, under Small Focus.
Set the large focus filament standby
current via the Standby Current (A)
dialog box, under Large Focus.
Set the small focus maximum filament
current via the Max Current (A) dialog
box, under Small Focus.
Set the large focus maximum filament
current via the Max Current (A) dialog
box, under Large Focus.
Ch # 901476-04
Rev. C