Control Console - CPI CMP 200 Series Service Manual

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

2
Installation
2.6.0
WIRING TO THE GENERATOR (Cont)
2.6.1

Control Console

Membrane Console
1.
Note the protective cover connected to the console cable. This is intended to protect the console
cable connectors during shipping and while routing the console cable during installation.
Disconnect the generator end of the cable (the end with the ferrite bead) from the protective
cover, and then route the end of the cable with the protective cover attached as required.
Remove and discard the protective cover when finished. After removing the protective cover,
inspect the console cable connectors for any damage. Please see figure 2-2 for an example of
such damage.
2.
Route the generator end of the console cable into the generator cabinet via the cable access slot
nearest to the generator control board. The cable must be routed as per figure 2-3. Connect the
generator end of the console cable to J3 on the generator control board.
3.
Connect the free end of the console cable to J8 at the rear of the console. Leave sufficient slack
in the cabling to the console to allow for future service and maintenance.
Touchscreen Console
1.
Route the generator end of the console cable into the generator cabinet via the cable access slot
nearest to the generator control board. The cable must be routed as per figure 2-3. Connect the
generator end of the console cable to J20 on the generator control board.
2.
Connect the free end of the console cable to GEN on the rear of the touchscreen console. Leave
sufficient slack in the cabling to the console to allow for future service and maintenance.
3.
Tighten the screw locks to fully secure the "D" connectors.
4.
Connect one end of the supplied ground wire to the touchscreen console ground screw, shown in
figure 2-4. Connect the other end of the ground wire to the ground connector near the main line
fuses in the generator (figure 2-7). Dress the ground wire away from high voltage areas in the
generator. This ground location may have other ground wires already connected; ensure that
these existing ground wires are not disconnected when making the X-ray tube ground connection.
Note that this ground wire is required for EMC compliance only. It has no safety impact.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 2-6
Figure 2-2: Console cable connector damage example
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-02
Rev. C