CPI Canada Inc.
3D.3.0 AEC CALIBRATION (TABLE BUCKY) Cont
400 speed film screen
Break Point
75 kVp knee pt.
55 kVp
50 kVp
65 kVp
110 kVp
130 kVp
85 kVp
95 kVp
800 speed film screen
Break Point
75 kVp knee pt.
55 kVp
50 kVp
65 kVp
110 kVp
130 kVp
85 kVp
95 kVp
NOTE: For SID's other than 40 in. (100 cm) multiply the mAs by the factor [new SID / 40 in. (100 cm)]
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
Service Manual
Absorber
mAs
20 cm H
O
4 mAs
2
15 cm H
O
6.3 mAs
2
15 cm H
O
10 mAs
2
15 cm H
O
2.5 mAs
2
25 cm H
O
1.6 mAs
2
25 cm H
O
1.25 mAs
2
20 cm H
O
2.5 mAs
2
20 cm H
O
1.25 mAs
2
Absorber
mAs
20 cm H
O
2 mAs
2
15 cm H
O
3.2 mAs
2
15 cm H
O
5 mAs
2
15 cm H
O
1.25 mAs
2
25 cm H
O
0.8 mAs
2
25 cm H
O
0.63 mAs
2
20 cm H
O
1.25 mAs
2
20 cm H
O
0.63 mAs
2
Table 3D-8: Target breakpoint calibration factors (Cont)
Ch # 901476-05
Rev. A
Generator mA
Generator BUT mAs
200 mA
200 mAs (MAX)
200 mA
200 mAs (MAX)
200 mA
200 mAs (MAX)
200 mA
200 mAs (MAX)
200 mA
200 mAs (MAX)
200 mA
200 mAs (MAX)
200 mA
200 mAs (MAX)
200 mA
200 mAs (MAX)
Generator mA
Generator BUT mAs
100 mA
120 mAs (MAX)
100 mA
120 mAs (MAX)
100 mA
120 mAs (MAX)
100 mA
120 mAs (MAX)
100 mA
120 mAs (MAX)
100 mA
120 mAs (MAX)
100 mA
120 mAs (MAX)
100 mA
120 mAs (MAX)
AEC Calibration