CPI CMP 200 Series Service Manual page 149

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.11.4 DAP Calculation Worksheet (Cont)
STEP
1.
Measured dose:
Convert mR to mGy if necessary by multiplying mR X
0.00873. Example 23.3 mR X 0.00873 = 0.203 mGy.
2.
Measure and record the exposed area of the film (Length
X Width).
Calculate the exposed area in cm
3.
Multiply the dose in mGy (step 1) X the area in cm
2). This will yield the actual dose-area product, in
2
mGycm
.
4.
Record the DAP, in mGycm
5.
Calculate the percentage error: Refer to the example at
the end of this section.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-80
ACTION
2
(length X width).
2
(step
2
, as displayed on the console.
Table 3C-4: DAP worksheets
CMP 200
RESULT
_____________ mR
_____________ mGy
_____________ Length (cm)
_____________ Width (cm)
2
_____________ Area (cm
_____________
2
(mGycm
)
_____________
2
(mGycm
)
_____________ % error
®
X-Ray Generator Service Manual
CPI Canada Inc.
)
DAP
DAP
Ch # 901476-04
Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cmp 200 dr series

Table of Contents