EMC (EN 60601-1-2:2001/IEC 60601-1-2:2001)
Guidance and manufacturer's declaration – electromagnetic emissions
The VZW2556 series of X-ray generators is intended for use in the electromagnetic environment specified
below. The customer or the user of the VZW2556 series should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
NOTE It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be
verified to assure that they meet the minimum specifications.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Compliance
Group 1
The VZW2556 series of X-ray generators use RF energy only for
their internal functions. Therefore, the RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
Class A
The VZW2556 series of X-ray generators must be used only in a
(The VZW2556
shielded location with a minimum RF shielding effectiveness and,
series of X-ray
for each cable that exits the shielded location, a minimum RF filter
generators in
attenuation of 40dB from 30 MHz to 230 MHz and 47dB from 230
combination with
MHz to 1 GHz. (The minimum at 30 MHz is 40dB and the
shielded location)
minimum at 230 MHz is 47dB).
Not Applicable
The VZW2556 series is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
Not Applicable
domestic purposes.
Electromagnetic environment - guidance
Ch # 901476-01
Rev. C
1
Pre-installation
Page 1-5