STATUS AND ERROR CODES

5.2.1

Operator Messages

These messages indicate the status of the generator.
MESSAGE
(membrane console)
DAP NOT READY
INITIALIZATION
SPINNING ROTOR
X-RAY ON
X-RAY READY
5.2.2

Limit Messages

These messages indicate that an exposure has been requested that exceeds one or more limits.
MESSAGE
(membrane console)
AEC DENSITY LIM.
ANODE HEAT WARN.
CAL LIMIT
DAP ACCUM. WARN
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
(touchscreen console)
DAP not ready
The optional DAP is in its "warm up" state, and not ready to make DAP
measurements.
Displayed during power up sequence.
Displayed when prep state is active.
Displayed during an X-ray exposure.
Generator Ready
Indicates that the generator is ready to make an exposure.
Generator offline...
Indicates that the console is unable to communicate with the generator.
Ensure that the generator is switched on, and that the console cable is properly
connected to the console.
Consult your service representative if this does not solve the problem.
MESSAGE
(touchscreen console)
Generator AEC density
Requested density not programmed.
limit
Anode warning level
Anode has exceeded programmed warning
exceeded
level.
Calibration limit: selected
Requested parameter not calibrated.
parameter not calibrated.
DAP accumulation
The accumulated DAP value exceeds the
Warning.
programmed DAP limit.
Ch # 901476-07
DESCRIPTION
PROBLEM
Rev. B
5
Troubleshooting
ACTION
Select another density or
program requested density
step.
Wait for anode to cool.
Recalibrate X-ray tube or
select a calibrated parameter.
Reset the DAP.
Page 5-3