3C.10.2 DAP SETUP Menu Items (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
DAP
PRINT
RESET
TEST
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
FUNCTION
DESCRIPTION
(GenWare®)
MEMBRANE CONSOLE ONLY: Allows access to a submenu used to reset the DAP display, test the DAP device,
and print a test label. This selection is available in the DAP SETUP menu when the DAP function is ON.
This function is
Prints a DAP label. The printer will print the date and time near the top of the label, and the
accumulated Dose-Area Product (mGycm
not available in
retrieved from the generator at the time the label is printed. Several headings are also printed
GenWare®
on the labels; the corresponding patient information must be manually entered.
Resets the DAP display to zero.
Reset
Tests the DAP circuits by counting the number of test pulses (refer to TEST VALUE 1, above).
Test
A pass / fail message will be presented after the DAP test, and the actual number of test pulses
counted will be displayed.
Ch # 901476-04
2
) near the middle of the label. This information is
Rev. C
Interfacing and Programming
Page 3C-73