3C.7.3 Receptor Setup (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
THE FOLLOWING SELECTIONS ARE ONLY AVAILABLE IF DEFAULTS WAS ENABLED AS PREVIOUSLY DESCRIBED.
TECHNIQUE
FOCUS
FILM SCREEN
LEFT FIELD
CENTER FIELD
RIGHT FIELD
KV
MA
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
FUNCTION
DESCRIPTION
(GenWare®)
Defines which technique will be defaulted to when a receptor is selected.
Technique
Membrane console:
MA:
MAS:
AEC
GenWare®:
ms:
mAs:
AEC
Focus
Defines which focus will be defaulted to when a receptor is selected. Options are SMALL or
LARGE.
Defines which film screen will be defaulted to when a receptor is selected and AEC enabled.
Film Screen
Options are film screen 1, 2, or 3 (membrane console) or I, II, or III GenWare®.
Fields
Defines which field(s) will be defaulted to when a receptor is selected.
Membrane console:
YES:
NO:
GenWare®:
Refer to the graphic under Fields:
A black (blue for Touchscreen GenWare®) field select rectangle = field not selected, a green
field select rectangle = field selected.
Voltage (kV)
Selects the default kV.
Current (mA)
Selects the default mA.
Ch # 901476-04
Defaults to mA/ms mode.
Defaults to mAs mode.
Defaults to AEC mode.
Defaults to mA/ms mode.
Defaults to mAs mode.
Defaults to AEC mode.
The selected field will be selected.
The selected field will not be selected.
Rev. C
Interfacing and Programming
Page 3C-49