CPI CMP 200 Series Service Manual page 181

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.13.10 Touchscreen Console Firmware Upgrade / Restore (Cont)
Step
5.
The progress of the installation process will be shown on the Updating Touch Screen window.
Note: In the event the upgrade should fail, the details of the error will be displayed in red at the bottom of the screen
and a pop-up window will display asking if you would like to retry or abort the upgrade. The cause of the failure
should be determined before retrying the upgrade.
Once the installation process is finished a pop-up window will display stating: It is very important that you do not disturb this
6.
process (see figure 3C-37). The flash memory is being erased and reloaded with the upgraded firmware.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-112
Figure 3C-36: Checking for Available Updates.
Action
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C