3D.3.0 AEC CALIBRATION (TABLE BUCKY) Cont
Figure 3D-6a - PC GenWare
Use these steps to perform the table Bucky AEC calibration.
Step
IF THE AEC BOARD BEING CALIBRATED HAS SHORT AEC TIME COMPENSATION POTENTIOMETERS, THE SHORT
EXPOSURE TIME COMPENSATION MUST FIRST BE DISABLED.
TO DO THIS, ADJUST ALL SHORT AEC EXPOSURE TIME COMPENSATION POTS TO ZERO BY TURNING EACH OF
THESE POTENTIOMETERS FULLY CLOCKWISE. THESE ARE MULTI-TURN POTENTIOMETERS, AND MUST BE TURNED
BY AS MUCH AS 25 TURNS TO REACH THE ZERO-OHMS LIMIT.
FAILURE TO PRESET THESE POTS WILL RESULT IN DIFFICULTY IN PERFORMING AEC CALIBRATION.
1.
Set up the X-ray tube stand as shown in figure 3D-5.
2.
Align the tube stand and table Bucky such that the central ray is centered relative to the image receptor.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
Service Manual
®
AEC Setup & Calibration window, Film Screen Calibration tab
Ch # 901476-05
Rev. A
Figure 3D-6b - Touchscreen GenWare
Action
3D
AEC Calibration
®
Page 3D-14