CPI CMP 200 Series Service Manual page 173

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.13.6 Touchscreen Setup (Cont)
Step
4.
Press OK to continue; CANCEL cancels the changes.
5.
A pop-up window will display asking if you wish to save the changes. Press YES to save the changes. Press NO to return to the
system utilities menu; CANCEL cancels the changes.
BE SURE TO RECORD THE NEW SERVICE PASSWORD BEFORE CHANGING THE FACTORY-DEFAULT PASSWORD. IF
THE NEW PASSWORD IS SUBSEQUENTLY LOST, CONSULT THE FACTORY OR RE-INSTALL THE TOUCHSCREEN
SOFTWARE IN ORDER TO RESTORE THE FACTORY-DEFAULT SERVICE PASSWORD.
To change the service password, press the ... button to the right of the top service password line. A password window will pop
6.
up. Enter the new password.
Press Clear to cancel an incorrect entry.
Press Cancel to close the password pop-up window.
Press Accept to accept the new password.
7.
Repeat the previous step to enter the new password on the second service password line.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-104
Figure 3C-32: Touchscreen setup window, Passwords tab
CMP 200
Action
®
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C