CPI CMP 200 Series Service Manual page 121

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.7.3 Receptor Setup (Cont)
Step
Action (membrane console)
10.
Press >>.
11.
Select SF/LF SWITCH. Toggle the button
to select AUTO or MAN.
12.
13.
Select AEC BACKUP. Toggle the button
to select FIXED, MAS, or MS.
14.
Select AEC BACKUP MAS. Use the + or
– buttons to select the maximum backup
mAs.
15.
Select AEC BACKUP MS. Use the + or –
buttons to select the maximum backup
ms.
16.
Select AEC CHANNEL. Use the + or –
buttons to assign the desired AEC
channel to the selected receptor, or to
disable AEC operation on that receptor.
17.
Press >>.
18.
Select RAD HANG. Use the + or –
buttons
to
hangover time.
The following steps only apply if MEMORY in step 6 was set to DEF / default.
19.
Select DEFAULTS.
Select TECHNIQUE. Toggle the button to
select MA, MAS, or AEC.
20.
Select FOCUS. Toggle the button to
select SMALL or LARGE.
21.
Select FILM SCREEN. Toggle the button
to select 1, 2, or 3.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-52
This function is not available in GenWare®
Select the AEC tab.
Under AEC Mode, select fixed, mAs, or
ms.
Select the maximum backup mAs via the
AEC Back-Up mAs dialog box.
Select the maximum backup ms via the
AEC Back-Up ms dialog box.
Select the AEC channel to be assigned to
the selected receptor, or disable AEC
operation on that receptor via the AEC
Channel dialog box.
Select the Receptor Properties tab.
Select the desired rad hangover time via
select
the
desired
rad
the Rad Hangover (s) dialog box.
Select the Receptor Defaults tab.
Under Technique, select ms, mAs, or
AEC.
Under Focus, select small or large.
Under Film Screen, select I, II, or III.
Action (PC GenWare®)
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Action (Touchscreen GenWare®)
Select the AEC tab.
Under AEC Back-up Mode, select
Fixed, mAs, or ms.
Select the maximum backup mAs via the
AEC Backup mAs dialog box.
Select the maximum backup ms via the
AEC Backup ms dialog box.
Select the AEC channel to be assigned
to the selected receptor, or disable AEC
operation on that receptor via the AEC
Channel dialog box.
Select the Receptor Properties tab.
Select the desired rad hangover time via
the Rad Hangover (s) dialog box.
Select the Receptor Defaults tab.
Under Technique, select ms, mAs, or
AEC.
Under Focus, select Small or Large.
Micro focus is not available at this time.
Under Film Screen, select I, II, or III.
Ch # 901476-04