3C.7.5 AEC Setup (Cont)
Note the following regarding GenWare® figure 3C-18:
The AEC Channel display near the top left side of the AEC Setup & Calibration window shows the selected AEC channel.
The Progr. Receptors display to the right of the AEC channel display shows which receptors are programmed for the
selected AEC channel.
Definitions of AEC SETUP menu items applicable to the initial AEC setup.
FUNCTION
(MEMBRANE
CONSOLE)
CHANNEL
LEFT FIELD
CENTER FIELD
RIGHT FIELD
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
FUNCTION
DESCRIPTION
(GenWare®)
Channel Number
Selects the AEC channel to be programmed.
A graphic
Enables or disables the ability to select the left, center, or right AEC fields.
depicting the AEC
Membrane console:
chamber fields.
YES:
NO:
GenWare®:
Refer to the graphic under Fields:
A black field select rectangle = disabled, a green field select rectangle = enabled.
Ch # 901476-04
The selected field is enabled.
The selected field is disabled.
Rev. C
Interfacing and Programming
Page 3C-63