X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
2.6.1
Control Console (Cont)
5.
Optional: Connect customer-supplied speakers (minimum 8 ohms) to LO (3.5 mm stereo jack) on
the rear of the touchscreen console (do not use externally amplified speakers). Use cable ties to
secure the speaker cable to the console cable for support. Pins 1 and 2 (left and right) on the 3.5
mm male stereo jack must be shorted together on the cable that plugs into the touchscreen
console.
Please note that connecting external speakers will disable one of the on-board touchscreen
speakers.
Connecting the touchscreen console cable to any connector other than J20 on the
Note:
generator control board may cause damage to the generator. Verify that the console cable
has been connected correctly prior to powering up the generator.
Mini-Console (optional)
1.
Route the generator end of the console cable into the generator cabinet via the cable access slot
nearest to the generator control board. The cable must be routed as per figure 2-3. Connect the
generator end of the console cable to J19 on the generator control board.
2.
Connect the free end of the console cable to the 15-pin male connector that is attached to the
short piece of console cable connected to the mini-console. Tighten the screw locks to fully
secure the "D" connectors on the console cable.
3.
An extra cable (9-pin 'D' to RJ45) is included with the optional mini-console. This must be
connected from the standard serial port (9-pin 'D' connector) on the imaging computer to
J3 on the generator control board on any DR systems that require serial communication
with the generator.
Note:
Connecting the mini-console cable to any connector other than J19 on the generator
control board may cause damage to the generator. Verify that the console cable has been
connected correctly prior to powering up the generator.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Ch # 901476-02
Rev. C
2
Installation
Page 2-7