8
Spares
8.2.0
SPARE PARTS LIST (Cont)
NOTE:
1.
Refer to table 8-1 for the part number for the H.V. auxiliary board. Four different part numbers are
used, depending on the application:
STATOR TYPE
"R" type stator
GE 23/23 Ω stator
Units with the dual speed starter option will be fitted with H.V. auxiliary board 739442-00.
Before installing a replacement H.V. auxiliary board, jumpers must be properly installed on this board
in order to select the Bucky voltage (24 VDC, 110 VAC, or 220 VAC). After determining the Bucky
drive voltage requirement, connect the appropriate pair of jumpers in accordance with table 8-2
The jumpers that select the Bucky voltage must be set per table 8-2.
BUCKY
OUTPUT
24 VDC
110 VAC
220 VAC
After determining the required rotor boost voltage, connect a wire jumper between the desired tabs on
the boards
ROTOR BOOST
VOLTAGE
120 VAC
240 VAC
*
A complete set of jumpers is shipped with each new generator. Three of the jumpers are installed on
the H.V. auxiliary board, and the remaining jumpers are in a bag attached to the lip on the inside of
the cable access slot above the HT tank.
If the Bucky or low-speed starter boost voltage needs to be changed on an existing H.V. auxiliary
board, use the existing jumper(s) if they are of the correct length and have the proper connector to fit
the tab(s) on the board. Otherwise, select the shortest wires from the spare jumper set that will
connect between the desired tabs on the board and that have the proper connectors on the jumpers.
For spares boards, the jumpers will need to be removed from the "old" board and reused.
Additionally, JW1 must be set per table 8-3 before installing a replacement H.V. auxiliary board.
UNIT TYPE
1 phase
3 phase
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 8-6
H.V. AUXILIARY BOARD JUMPERS
(units with isolated 110 / 220 VAC, using auxiliary transformer 739446)
E11-E5, E7-E8
E9-E5, E7-E8
E12-E5, E7-E8
*
.
H.V. AUXILIARY BOARD JUMPERS
E14-E15
E16-E15
JW1 CONFIGURATION: H.V. AUXILIARY BOARD
OPEN (jumper pins 2-3)
CLOSED (jumper pins 1-2)
®
CMP 200
X-Ray Generator Service Manual
TABLE 8-1
208 / 230 V GENERATORS
739445-00
739445-01
TABLE 8.2
TABLE 8-3
TABLE 8-4
CPI Canada Inc.
400 / 480 V GENERATORS
739442-00
739442-01
Ch # 901476-10
*
.
Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cmp 200 dr series

Table of Contents