CPI CMP 200 Series Service Manual page 102

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
3C.7.1 Tube Selection / Tube Setup (Cont)
Definitions of TUBE SELECTION menu items.
FUNCTION
(MEMBRANE
CONSOLE)
TUBE SPEED
MAX SF KW LS
MAX LF KW LS
MAX KV
SF STANDBY
LF STANDBY
SF MAX
LF MAX
FIL BOOST
FIL PREHEAT
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
FUNCTION
DESCRIPTION
(GenWare®)
Tube Speed
Selects the dual speed starter mode (dual speed starter option only).
LOW:
HIGH:
DUAL:
Sets the maximum small focus low speed kW limit (@ 100 ms *).
Low Speed: Max
SF Power (kW)
Sets the maximum large focus low speed kW limit (@ 100 ms *).
Low Speed: Max
LF Power (kW)
Sets the maximum kV allowed.
Max. Tube
Voltage (kV)
Sets the small focus standby filament current. The required value should be obtained from the
Small Focus:
X-ray tube data sheets.
Standby Current
(A)
As above but for large focus.
Large Focus:
Standby Current
(A)
Small Focus: Max
Sets the small focus maximum filament current.
Current (A)
Large Focus: Max
As above but for large focus.
Current (A)
Filament Boost
Sets the filament rapid boost duration in order to quickly raise the filament temperature.
Time (ms)
The time that the filament is held at the required emission level before an exposure is permitted.
Filament Preheat
Time (ms)
Ch # 901476-04
Low speed mode only is enabled.
High speed mode only is enabled.
Low and high-speed modes are both enabled. The generator will determine when to
switch modes, based on exposure parameters and tube rating.
Rev. C
Interfacing and Programming
Page 3C-33