2.2 Outputs - CPI CMP 200 Series Service Manual

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.2.1 Inputs (Cont)
H.V. AUXILIARY BOARD
J2-5 (-)
J2-6 (+)
J4-5 (-)
J4-6 (+)
J4-9 (-)
J4-10 (+)
J2-1 (-)
J2-2 (+)
J2-3 (-)
J2-4 (+)
*
These inputs are meant for dry contacts only. Do not apply any voltage source to these inputs.
3C.2.2 Outputs
Table 3C-2 shows the Bucky and auxiliary power outputs from the generator. Refer also to figure 3C-1.
H.V. AUXILIARY BOARD
J2-10
J2-9
J2-8
J2-7
J4-4
J4-3
J4-2
J4-1
J11-3
J11-4
J11-5
J11-6
J11-1 (-)
J11-2 (+)
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-4
B2:
Bucky 1 ready. A contact closure from pin 5 to 6 indicates
B1:
BUCKY 1 READY.
B2:
Bucky 2 ready. A contact closure from pin 5 to 6 indicates
B1:
BUCKY 2 READY.
Door interlock. A contact closure from pin 9 to 10 indicates that the
room door is closed.
Interlock 2. A contact closure from pin 1 to 2 indicates that interlock 2
*
is closed.
This is the tomo exposure input for any image receptor that is
programmed for tomo operation. The tomo start / stop command is
generated by the tomo system. The generator will only terminate the
tomo exposure if the tomo backup time is exceeded.
Interlock 1. A contact closure from pin 3 to 4 indicates that interlock 1
*
is closed.
Table 3C-1: Generator inputs
B4: Bucky 1 return.
B6: Ground (Bucky 1).
B8: 24 VDC, 110 / 220 VAC out (Bucky 1). See note below.
B3: 24 VDC, 110 / 220 VAC out (Bucky 1 start). See note below.
B4: Tomo / Bucky 2 return.
B6: Ground (Tomo / Bucky 2).
B8: 24 VDC, 110 / 220 VAC out (Tomo / Bucky 2). See note
below.
B3: 24 VDC, 110 / 220 VAC out (Tomo / Bucky 2 start). See note
below.
Room lights. The generator supplies a dry contact closure when
the room light is to be activated.
24 VAC @ 150 watts output for customer use.
24 VDC @ 45 watts output for customer use (unswitched).
Table 3C-2: Generator outputs
®
CMP 200
X-Ray Generator Service Manual
DEFINITION
*
*
*
DEFINITION
CPI Canada Inc.
*
Ch # 901476-04
Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cmp 200 dr series

Table of Contents