CPI CMP 200 Series Service Manual page 264

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
5.2.3
Error Messages (Cont)
ERROR
MESSAGE
CODE
(membrane console)
E082
KV OVER VOLTAGE
E083
ANODE MA FAULT
E084
CATHODE MA FAULT
E085
ROTOR MAIN FAULT
E086
ROTOR SHFT FAULT
E100
CAL-MAX MA ERR.
E101
CAL-DATA LIMIT
E102
CAL-MAX FIL ERR
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
PROBLEM
(touch screen console)
kV Over Voltage
The output kV exceeded the
maximum allowed kV limit.
Anode mA Fault
The anode current exceeded the
maximum allowed limit.
Cathode mA Fault
The cathode current exceeded the
maximum allowed limit.
Rotor Main Fault
Indicates that insufficient current was
sensed in the stator main winding.
Rotor Shift Fault
Indicates that insufficient current was
sensed in the stator shift winding.
Calibration error –
Maximum mA has been exceeded
maximum mA exceeded
during auto calibration.
Calibration error –
Auto calibration has exceeded data
calibration data table
table length due to an excessive
exceeded
number of exposures.
Calibration error –
Maximum filament current for the
maximum filament
selected focus has been reached.
current exceeded
Ch # 901476-07
Rev. B
5
Troubleshooting
ACTION
1. Auto calibrate tube (per
TUBE CALIBRATION in
*
chapter 3C)
.
2. Consult product support.
1. Check for arcing of the X-
ray tube.
2. Check for arcing in the HT
tank; if failure of HT tank is
suspected, contact product
support.
1. Check for arcing of the X-
ray tube.
2. Check for arcing in the HT
tank; if failure of HT tank is
suspected, contact product
support.
1. Check the X-ray tube stator
wiring.
2. Check the rotor board.
1. Check the X-ray tube stator
wiring.
2. Check the rotor board.
Repeat auto calibration and/or
decrease standby current.
1. Check to see if the filament
standby current is too low.
2. Retry auto calibration.
1. Check to see if the
maximum filament current
limit can be increased.
2. Retry auto calibration.
Page 5-11