3C.7.2 Generator Limits (Cont)
Use these steps to set the generator limits. Refer to the definitions in the previous table.
Step
Action (membrane console)
1.
From the GEN CONFIGURATION menu,
select GENERATOR LIMITS.
17.
Select MAX KW. Use the + or – buttons
to set the maximum kW.
18.
Select MAX MA. Use the + or – buttons to
set the maximum mA.
2.
Select MIN MA. Use the + or – buttons to
set the minimum mA.
3.
Select MAX MAS. Use the + or – buttons
to set the maximum mAs.
4.
Select ROTOR BOOST. Use the + or –
buttons to set the low speed starter boost
time.
Press EXIT to return to the GEN
5.
CONFIGURATION menu.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Action (PC GenWare®)
Select Generator Limits from the Setup
menu, or use the generator limits button
on the GenWare® toolbar.
Set the maximum kW via the Max.
Generator Power (kW) dialog box.
Set the maximum mA via the Max.
Generator Current (mA) dialog box.
Set the minimum mA via the Min.
Generator Current (mA) dialog box.
Set the maximum mAs via the Max.
Generator Current-Time (mAs) dialog
box.
This function is not available in
GenWare®
Ch # 901476-04
Rev. C
Interfacing and Programming
Action (Touchscreen GenWare®)
Press the
button to access the
Generator Limits utility.
Set the maximum kW via the Max
Generator Power (kW) dialog box.
Set the maximum mA via the Max
Generator Current (mA) dialog box.
Set the minimum mA via the Min
Generator Current (mA) dialog box.
Set the maximum mAs via the Max
Generator Current-Time (mAs) dialog
box.
This function is not available in
GenWare®
Page 3C-41