5.2.3
Error Messages (Cont)
ERROR
MESSAGE
CODE
(membrane console)
E055
NO FIELDS ACTIVE
E056
NO TUBE SELECTED
E057
AEC STOP ERROR
E058
CONSOLE BUT ERR.
E060
EXP. KV HIGH
E061
EXP. KV LOW
E063
FACTORY DEFAULTS
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
PROBLEM
(touchscreen console)
No fields selected in
AEC enabled but no fields are
AEC mode
selected.
Receptor Disabled
All Receptors have no X-ray tube
programmed.
AEC stop signal In
AEC stop signal (P.T. stop signal) is
wrong state
active low indicating exposure is
finished during prep state.
Console back-up timer
Console has detected exposure
exceeded backup time and
terminated exposure.
High kV error
kV exceeds high kV tolerance level.
Low kV error
kV exceeds low kV tolerance level.
Factory defaults set
S3-1 on the generator control board
is set to the LOAD DEFAULTS
position.
Ch # 901476-07
Rev. B
5
Troubleshooting
ACTION
Select AEC field(s).
Program receptor(s) with tube
number.
1. Check that the P.T. ramp
does not exceed the P.T.
reference during prep state.
2. Check AEC device for
proper operation.
Call product support.
1. Check the output of the kV
reference DAC on the
generator CPU board.
2. Measure the output of the
generator with a dynalyzer
or a non-invasive kVp
meter.
1. Check the output of the kV
reference DAC on the
generator CPU board.
2. Measure the output of the
generator with a dynalyzer
or a non-invasive kVp
meter.
Set S3-1 to the NORMAL
position. The generator will not
exit the initialization phase
until this is done.
Page 5-9