CPI Canada Inc.
5.2.3
Error Messages (Cont)
ERROR
MESSAGE
CODE
(membrane console)
E065
TOMO DEVICE ERR.
E067
PS DUTY LIMIT
E071
DAP DOSE OVERFLW
E072
DAP DEVICE ERR.
E073
DAP DATA ERROR
E074
INVERTER 1 ERROR
E075
INVERTER 2 ERROR
E077
RES. CIRCUIT ERR
E078
BUCKY1 INTERLOCK
E079
BUCKY2 INTERLOCK
E080
INTERLOCK 1 OPEN
E081
INTERLOCK 2 OPEN
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
PROBLEM
(touchscreen console)
Tomo device error.
The tomo start signal was not
received within 30 seconds of
pressing the X-ray switch on the
generator.
Power supply duty cycle
Power supply duty cycle limit
limit exceeded
exceeded.
The accumulated DAP value
DAP overflow error
exceeds the display limit.
The DAP device is not functional.
DAP device error
The DAP configuration data is
DAP data error
corrupted.
Power circuit failure.
Inverter 1 Error
Power circuit failure.
Inverter 2 Error
HT primary overcurrent detected.
Resonant Circuit Error
Bucky 1 Interlock error
Indicates that Bucky 1 is not ready.
Bucky 2 Interlock error
Indicates that Bucky 2 is not ready.
Interlock 1 Open
Indicates that the interlock 1 is open.
Interlock 2 Open
Indicates that the interlock 2 is open.
Ch # 901476-07
Rev. B
5
Troubleshooting
ACTION
1. Check the tomo
connections to the
generator.
2. Check the tomo system.
Re-evaluate technique factors.
Allow generator to cool if
possible. Exposures are
inhibited until the generator
has cooled sufficiently.
Reset the DAP.
1. Check the DAP wiring.
2. Check the DAP interface
board.
Reset factory defaults.
Contact product support.
Contact product support.
1. Check for arcing of the X-
ray tube.
2. Check for arcing in the HT
tank; if failure of HT tank is
suspected, contact product
support.
Check Bucky 1.
Check Bucky 2.
Check the circuits connected
to interlock 1. Refer to INPUTS
in chapter 3C.
Check the circuits connected
to interlock 2. Refer to INPUTS
in chapter 3C.
Page 5-10