13.7 Touchscreen Calibration - CPI CMP 200 Series Service Manual

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
3C.13.6 Touchscreen Setup (Cont)
Step
8.
Press OK to continue; CANCEL cancels the changes.
9.
A pop-up window will display asking if you wish to save the changes. Press YES to save the changes. Press NO to return to the
system utilities menu; CANCEL cancels the changes.
3C.13.7 Touchscreen Calibration
The touchscreen calibration function should be performed if touching the center of a buttons does not activate that function, i.e. if the touch-
sensitive area appears displaced from the center of the corresponding icon.
Use these steps to perform the touchscreen calibration.
Step
1.
From the system utilities menu, press Touch Screen Calibration.
2.
A white screen with a cross-hairs target will appear in the top left corner of the display.
Press the center of the target at the top left of the screen. This will set up the touch sensitive area in that quadrant.
3.
Repeat the previous step for the four remaining targets.
4.
Another white screen will appear with an X enclosed by a box in the top left corner and a 10 second timer in the center of the
5.
display.
Press the center of the target at the top left of the screen. This will verify the touch sensitive area in that quadrant.
6.
Repeat the previous step for the four remaining targets.
7.
If the calibration procedure was done properly you will be taken to the System Utilities menu.
8.
If the calibration procedure was not done properly the calibration sequence will restart from the beginning allowing you to
redo the touchscreen calibration.
NOTE:
IF THE TOUCHSCREEN CALIBRATION IS OUT SIGNIFICANTLY, CONNECT A USB (PC) KEYBOARD TO ONE OF THE USBA
PORTS ON THE REAR OF THE TOUCHSCREEN CONSOLE, POWER UP THE UNIT AND WHEN THE MAIN MENU IS
DISPLAYED PRESS ALT-C TO INITIATE THE CALIBRATION PROCEDURE. FOLLOW THE STEPS ON THE PREVIOUS PAGE
TO COMPLETE THE TOUCHSCREEN CALIBRATION.
CPI RECOMMENDS NO OTHER BUTTONS BE PRESSED ON THE USB KEYBOARD AS THIS MAY INADVERTENTLY CAUSE
CORRUPTION OF THE TOUCHSCREEN SOFTWARE.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Action
Action
Ch # 901476-04
Rev. C
Interfacing and Programming
Page 3C-105