3C
Interfacing and Programming
3C.13.10 Touchscreen Console Firmware Upgrade / Restore (Cont)
Step
The progress of the installation will be shown on the Updating Touch Screen window.
10.
NOTE: IN THE EVENT THE UPGRADE SHOULD FAIL, THE DETAILS OF THE ERROR WILL BE DISPLAYED IN RED AT
THE BOTTOM OF THE SCREEN AND A POP-UP WINDOW WILL DISPLAY ASKING IF YOU WOULD LIKE TO
RETRY OR ABORT THE UPGRADE. THE CAUSE OF THE FAILURE SHOULD BE DETERMINED BEFORE
RETRYING THE UPGRADE.
Once the installation process is finished, a pop-up window will display stating: It is very important that you do not disturb this
11.
process (see figure 3C-44). The flash memory is being erased and reloaded with the upgraded firmware.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-120
Figure 3C-44: Erasing Flash memory pop-up window
®
CMP 200
Action
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C