13.1 Accessing The Utilities Menu - CPI CMP 200 Series Service Manual

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.13.1 Accessing the Utilities Menu
Use these steps to access the systems utilities functions.
Step
1.
From the main console menu (figure 3C-22), press System Utilities. A pop-up window will be displayed (figure 3C-23),
requesting a password.
2.
Press 1, 9, 7, 3 in sequence to continue. This is the factory-default service password, and allows access to all of the functions
listed above.
Press Clear to cancel an incorrect password.
Press Cancel to return to the main menu.
Press Accept to access the system utilities menu. After a brief delay, the system utilities menu (figure 3C-24) will be
displayed. The message Access Denied indicates that an incorrect password was used. The factory-default password
may be changed by a service engineer as described later in this supplement. If this was done, the password defined
above will not allow access to the system utilities menu.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-84
Action
Figure 3C-22: Main menu
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C