Pre-installation
1.1.0

INTRODUCTION

This chapter summarizes the main features of the CMP 200® and CMP 200® DR X-ray generators
(performance, regulatory and compatibility). Safety information is provided, along with environmental,
room, and installation requirements. This chapter concludes with a pre-installation checklist and a
diagram showing the major component layout.
The information in this chapter is provided in order for the installer to be able to plan the site
layout prior to installation of the generator.
1.1.1

Generator Description

The CMP 200® X-ray generator is intended for use in film-based stationary radiographic X-ray systems,
including tomography. The CMP 200® DR X-ray generator adds a digital interface for digital radiography
(DR) equipment. The CMP 200® X-ray generator consists of a main power cabinet and an optional
membrane control console. The CMP 200® DR X-ray generator consists of a main power cabinet and an
optional membrane, touchscreen or mini-console (used with digital interface). The main power cabinet
contains the HT tank and control circuits, the filament drivers, a low speed starter (optional dual speed
starter on some models of CMP 200® DR), and interface connections to the room equipment.
The control console allows the operator to select the technique factors, image receptors, etc.,
and to initiate an X-ray exposure.
1.1.2

Features

The following are the main features of and the options available for the generator:
Integral low speed starter, compatible with X-ray tubes with type "R" stator. Optional compatibility with
GE 23/23 Ω equal impedance "E" stator.
Optional dual-speed starter on some models of CMP 200® DR (not available on 208 / 230 VAC, 1-
phase & 3-phase units), compatible with tube types listed in chapter 2.
Capable of interfacing with various DR imaging systems (CMP 200® DR only).
24 VDC, 110, or 220 VAC power source for Buckys.
24 VAC 150 watts power source for collimator lamp.
24 VDC 45 watts power source for system locks.
Optional AEC.
Optional DAP (Dose-Area Product).
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-2
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-01
Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cmp 200 dr series

Table of Contents