CPI Canada Inc.
22.
Select LEFT FIELD. Toggle the button to
select YES or NO.
3C.7.3 Receptor Setup (Cont)
Step
Action (membrane console)
23.
Repeat the previous step for the CENTER
and RIGHT fields.
24.
Press >>.
25.
Select KV, MA, MS, and DENSITY. Use
the + or – buttons to select the default kV,
mA, ms, and density, respectively.
26.
Press << and EXIT as required to the
GEN CONFIGURATION menu.
27.
Repeat steps 4 to 25 for the remaining
receptors. It is necessary to return to the
GEN
CONFIGURATION
programming each receptor to ensure
that the updated parameters are saved to
memory.
3C.7.4 I/O Configuration
The I/O CONFIGURATION function allows programming the states of the exposure for the inputs and the outputs on the room interface
board.
The I/O CONFIGURATION menus for the membrane console are shown below.
TUBE THERMAL SW.
DOOR ITLK:
EXIT
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Click the left field on the graphic under
Fields to select / deselect that field. Black
indicates that the field is not selected;
green indicates that the field is selected.
Repeat the previous step for the center
and right fields.
Select the default kV, mA, ms, and
density via the Voltage (kV), Current
(mA), Time (ms), and Density dialog
boxes, respectively.
menu
after
* INPUTS [sym] *
_ _ _ _
STANDBY STATE
_ _ _ _
Ch # 901476-04
Action (PC GenWare®)
>>
Rev. C
Interfacing and Programming
Click the left field on the graphic under
Fields to select / deselect that field.
Black indicates that the field is not
selected; green indicates that the field is
selected.
Action (Touchscreen GenWare®)
Repeat the previous step for the center
and right fields.
Select the default kV, mA, ms, and
density via the Voltage (kV), Current
(mA), Time (ms), and Density (s) dialog
boxes, respectively.