2.3 Inputs & Outputs (Simplified Schematic) - CPI CMP 200 Series Service Manual

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.2.3 Inputs & Outputs (simplified schematic)
BUCKY 1
OUTPUTS
BUCKY 1
READY *
INTERLOCK
INTERLOCK #2 /
TOMO EXPOSURE
INPUT *
TOMO / BUCKY 2
OUTPUTS
BUCKY 2
READY *
DOOR
INTERLOCK *
24 VAC OUT
FOR COLLIMATOR
LAMP
ROOM
LIGHTS
24 VDC OUT
FOR TUBE
STAND LOCKS
Figure 3C-1: J2, J4, and J11 inputs and outputs (H.V. auxiliary board)
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-6
J2
10
BUCKY RETURN
9
8
BUCKY 1 START
7
+24 VDC
6
5
4
#1 *
3
2
1
J4
BUCKY RETURN
4
3
2
BUCKY 2 START
1
+24 VDC
6
5
10
9
J11
F11
FROM 24 VAC
6
SOURCE
5
3
4
F12
FROM 24 VDC
+
2
1
*
REFER TO TABLE 3C-1 FOR REQUIRED INPUT LOGIC LEVELS
**
THE BUCKY OUTPUTS MAY BE CONFIGURED FOR +24 VDC, 110 VAC,
OR 220 VAC. REFER TO CHAPTER 8 FOR DETAILS.
®
CMP 200
X-Ray Generator Service Manual
** 24 VDC
110 / 220 VAC
R2
C3
K1
BUCKY 1 READY
+24 VDC
** 24 VDC
110 / 220 VAC
R15
C10
K3
BUCKY 2 READY
+24 VDC
+12 VDC
SOURCE
+12 VDC
K1
BUCKY 1 START
COMMAND
INTERLOCK 1
+24 VDC
INTERLOCK 2
+12 VDC
K3
BUCKY 2 START
COMMAND
DOOR INTERLOCK
K7
ROOM LIGHT
"ON" COMMAND
Ch # 901476-04
CPI Canada Inc.
FILE: CMP200_012B.CDR
Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for: