DIP SWITCH SETTINGS

Before continuing, verify the DIP switch settings on the generator control board. These switches have
been factory set but may have been readjusted, particularly if this generator is a re-install.
GENERATOR CONTROL BOARD:
Verify the settings on S3. Refer to the table below for the proper settings for this switch.
GENERATOR POWER
32 kW
40 kW
50 kW
65 kW
S3-1
DIP switches S3-2, S3-3, and S3-4 are not used.
*
Refer to Resetting Factory Defaults in chapter 6 for details regarding this function.

2.10.0 INITIAL RUN-UP

This section describes the procedure for auxiliary transformer line voltage tap selection, and for initial
power-on of the generator after it has first been installed.
PLEASE OBSERVE THE FOLLOWING POINTS REGARDING THE MAIN DISTRIBUTION
TRANSFORMER:
IF USING A DISTRIBUTION TRANSFORMER WITH AN ISOLATED SECONDARY, THE
SECONDARY WINDING MUST BE A WYE (STAR) CONFIGURATION WITH THE CENTER
POINT GROUND-REFERENCED. DO NOT USE A DELTA-CONFIGURED SECONDARY, AS
THERE IS NO GROUND REFERENCE IN THIS CONFIGURATION.
IF USING AN AUTOTRANSFORMER TYPE DISTRIBUTION TRANSFORMER, THE A.C.
INPUT TO THE TRANSFORMER MUST BE GROUND-REFERENCED.

2.10.1 Auxiliary Transformer Line Voltage Tap Selection

For 208 / 230 V generators, the line voltage taps on the auxiliary transformer must be checked before
powering up the generator.
For 400 / 480 V generators, the auxiliary transformer line voltage tap is factory set to match the line
voltage that was specified at the time of the order. If these units are to be operated from other than the
rated line voltage (i.e. if a 400 V generator is to be operated from 480 V mains), the line voltage tap on
the auxiliary transformer must be changed as described below.
For 208 / 230 VAC generators:
1.
Verify that the mains voltage and current capacity is correct for the generator installation per
Generator Power Requirements in chapter 1.
2.
Locate the auxiliary transformer inside the generator cabinet. Refer to chapter 1.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MAXIMUM mA
400 mA
500 mA
630 mA
800 mA
ON > LOAD FACTORY DEFAULTS
S3-8
S3-7
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
*
OFF > DEFAULTS DISABLED
Ch # 901476-02
Rev. C
Installation
S3-6
S3-5
OFF
ON
ON
OFF
ON
ON
OFF
ON
*
Page 2-31