Hand Switch (Optional); Stator Connections - CPI CMP 200 Series Service Manual

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

2
Installation
2.6.2

Hand Switch (Optional)

The optional hand switch, if ordered from CPI Canada Inc, is supplied pre-wired to a male 9 pin
subminiature "D" connector. This connects to J3 on the membrane console or to HS on the touchscreen
console. A male 9 pin subminiature "D" connector will need to be provided by the installer if the CPI
supplied hand switch is not used.
Pin assignments are as follows:
PIN NUMBER
1
2
3
4
5
6
7
8
9
2.6.3

Stator Connections

Low-speed Starter
Refer to figure 2-5 for the X-ray tube stator and thermal switch connections for the low speed starter.
1.
Route the X-ray tube stator cable to the stator terminal block J7 on the H.V. auxiliary board. This
is shown in figure 2-5.
USE OF A SHIELDED STATOR CABLE IS RECOMMENDED. THE SHIELD FOR THE
STATOR CABLE MUST BE PROPERLY GROUNDED AT BOTH THE TUBE END AND THE
GENERATOR END OF THE CABLE. CONNECT THE GENERATOR END OF THE SHIELD TO
THE GROUND TERMINAL ON J7.
2.
Connect the X-ray tube stator and the thermal switch to J7 on the H.V. auxiliary board terminal
block as per the table below.
Figure 2-5: Stator / thermal switch connections on H.V. auxiliary board (low speed starter)
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 2-10
(J3 membrane console) / HS touchscreen console)
X-ray
No Connection
Prep
No Connection
Common (ground)
NOT USED
NOT USED
NOT USED
NOT USED
CMP200_STATOR
®
CMP 200
X-Ray Generator Service Manual
PIN CONNECTIONS
J7
6
5
CPI Canada Inc.
4
3
2
1
Ch # 901476-02
Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cmp 200 dr series

Table of Contents