CPI CMP 200 Series Service Manual page 128

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

CPI Canada Inc.
3C.7.4 I/O Configuration (Cont)
Use these steps for programming the I/O functions. Refer to the definitions in the previous table.
Step
Action (membrane console)
From the GEN CONFIGURATION menu,
1.
select I/O CONFIGURATION.
Select the first receptor to be
2.
programmed.
3.
Press the STATE button to select the first
state that can be programmed for the
TUBE THERMAL SW. input. This is the
STANDBY state.
Toggle the TUBE THERMAL SW. button
to select the desired logic level (low or
high) to disable or enable that input
during the standby state.
Repeat the previous step for each state.
4.
Repeat steps 3 and 4 for each input. Use
5.
the >> button to advance to INPUTS
menu 2.
Press >> to select the OUTPUTS menu.
6.
Repeat steps 3 and 4 to program each
7.
output.
When finished the I/O programming for
8.
the current receptor, exit to the GEN
CONFIGURATION menu.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Action (PC GenWare®)
From the Receptor Setup window, select
the Inputs tab.
Select the first receptor to be
programmed. Use the Receptor tabs on
the Receptor Setup window.
For the Tube Thermal Switch input,
identify the first state that can be
programmed. This is the Standby state.
Enable or disable the Tube Thermal
Switch input during the standby state by
checking (with an X), or unchecking the
Standby column to the right of Tube
Thermal Switch.
Repeat the previous step for each state.
Repeat steps 3 and 4 for each input.
Select the Outputs tab.
Repeat steps 3 and 4 to program each
output.
An unchecked state causes the relay
connected to that output to be de-
energized during the selected state. A
checked state (with an X) will result in the
relay being energized during the selected
state.
Select Apply to save the programming for
the current receptor.
Ch # 901476-04
Rev. C
Interfacing and Programming
Action (Touchscreen GenWare®)
From the Receptor Setup window, select
the Inputs tab.
Select the first receptor to be
programmed. Use the Receptor buttons
on the Receptor window.
For the Tube Thermal Switch input,
identify the first state that can be
programmed. This is the Standby state.
Enable or disable the Tube Thermal
Switch input during the standby state by
checking (with an X), or unchecking the
Standby column to the right of Tube
Thermal Switch.
Repeat the previous step for each state.
Repeat steps 3 and 4 for each input.
Select the Outputs tab.
Repeat steps 3 and 4 to program each
output.
An unchecked state causes the relay
connected to that output to be de-
energized during the selected state. A
checked state (with an X) will result in the
relay being energized during the selected
state.
Select Apply to save the programming for
the current receptor.
Page 3C-59