Interfacing and Programming
3C.13.9 Configuration Backup / Restore (Cont)
To back-up current receptor / auto-position symbols.
Step
3.
Press Backup.
One or two keyboard(s) will display on the screen as described under APR Editor.
4.
To restore backed-up receptor / auto-position symbols.
Step
5.
Select the desired backup file from the configuration backup and restore menu. The selection will be highlighted.
6.
Press Restore.
7.
A pop-up window will briefly display indicating the file-restoration status. The file restoration is finished when the pop-up window
closes.
To delete a receptor / auto-position symbols file.
Step
8.
Select the backup file to be deleted. The selection will be highlighted.
9.
Press Delete. A pop-up window will display asking if you are sure you want to delete this file. Press YES to delete the file; NO
cancels the deletion.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-108
Type in the name of the new backup file.
Press ENTER when finished. The console will return to the configuration backup / restore menu, and the
backed up file will be displayed in the Configuration Backup And Restore window.
Action
Action
Action
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C