3D.9.0 PRECALIBRATION SETUP
3D.9.1 AEC Setup Worksheet
Before continuing, it is suggested that a copy of the table below be filled in with all required information.
Refer to the example AEC setup worksheet on the next page.
FUNCTION
Film/Screen
Nominal optical density:
Grid ratio/SID:
Min - max kV range:
± Density steps
Density mAs
(dose) change
(per step)
Chamber type:
Regulatory AEC dose
requirements?
Is film processing maintained?
Assigned receptor name:
Are all cassettes similar?
Additional notes:
Additional notes:
ALL RECEPTORS MUST HAVE THE SAME NUMBER OF DENSITY STEPS AND THE SAME DENSITY DOSE
CHANGE PER STEP (DENSITY SETTINGS ARE COMMON TO ALL FILM SCREENS AND RECEPTORS).
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
Service Manual
RECEPTOR 1
1.
2.
3.
+%:
-%:
Table 3D-11: AEC setup worksheet
Ch # 901476-05
RECEPTOR 2
Rev. A
3D
AEC Calibration
RECEPTOR 3
Page 3D-37