Error Messages (Cont)
ERROR
MESSAGE
CODE
(membrane console)
E008
KV/MA FAULT
E009
PS NOT READY
E011
HIGH MA FAULT
E012
LOW MA FAULT
E013
MANUAL TERMIN.
E014
AEC BUT ERROR
E015
AEC BU MAS ERR.
E016
TOMO BUT ERROR
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
MESSAGE
PROBLEM
(touchscreen console)
kV / mA Fault
The generator has detected a fault in
the kV or mA output during an
exposure and immediately
terminated the exposure. This may
be caused by arcing in the X-ray
tube, arcing of the HV cables, or HT
tank.
Power Supply Not Ready
The generator is not ready to make
an exposure.
mA During Exposure
Generator CPU detected mA greater
Too High
than allowed tolerance.
mA During Exposure
Generator CPU detected mA less
Too Low
than allowed tolerance.
Manually Terminated
Operator released exposure switch
Exposure
during exposure.
AEC Back-up Timer -
AEC exposure exceeded allowed
Exposure Terminated
back up time.
AEC mAs Exceeded -
AEC exposure exceeded allowed
Exposure Terminated
back up mAs.
Tomo Back-up Timer –
Tomo exposure exceeded backup
Exposure Terminated
time.
Ch # 901476-07
Rev. B
5
Troubleshooting
ACTION
1. If arcing of the X-ray tube is
suspected, check condition
of tube. The X-ray tube
may be damaged or simply
require "seasoning". Refer
to chapter 6 for tube
seasoning procedure.
2. If failure of HT tank is
suspected, contact product
support.
Retry exposure.
. *
Recalibrate X-ray tube
. *
Recalibrate X-ray tube
1. Re-take exposure if
necessary.
2. Check for faulty switch
contacts or wiring.
1. Check exposure technique
settings.
2. Check that correct AEC
chamber is energized.
1. Check exposure technique
settings.
2. Check that correct AEC
chamber is energized.
1. Check exposure technique
settings.
2. Increase tomo backup time,
if necessary.
Page 5-6