LINEARITY

1.
Record two additional series of dose measurements for entry into tables 6 and 7:
For table 6, use settings per table 3 EXCEPT use an mA (or mAs) value adjacent to the mA
(or mAs) setting used in table 3.
For table 7, use settings per table 4 EXCEPT use an mA (or mAs) value adjacent to the mA
(or mAs) setting used in table 4.
Record the mAs in the header of tables 6 and 7 as per 3.0 step 3.
1.
Calculate and record the average dose K (kerma) in mGy for tables 6 and 7.
2.
Record the preselected mAs and the average dose values taken from tables 3 and 4, and from
tables 6 and 7, at the top of the next page.
3.
Using the appropriate mAs and K values, calculate X
4.
Calculate the coefficient of linearity, L, as per tables 8 and 9.
mAs = __________
EXP No.
1
2
3
4
5
Table 6: Linearity
mAs = __________
EXP No.
1
2
3
4
5
Table 7: Linearity
Use and disclosure is subject to the restrictions on the title page of this CPI document.
X-Ray Generator Service Manual Supplement
DOSE (mGy)
K = ________
DOSE (mGy)
K = ________
Reproducibility, Linearity, & HVL Testing
, X
, X
, and X
3
4
6
740917-00
Rev. C
in tables 8 and 9.
7
Page 7