3C.13.10 Touchscreen Console Firmware Upgrade / Restore (Cont)
Step
4.
A Checking for Available Updates screen will display. It contains two windows: Current Packets and Upgrade Packets (see
figure 3C-36). Select the desired packets from the Upgrade Packets window. Packets that require upgrading will already be
selected. Press Next.
Note: If no upgrade is available for the U-boot packet, it will not be available for selection. This is done to prevent
software corruption should the touchscreen console inadvertently lose power.
Note: Any Upgrade Packet that is of an earlier version than the one already installed will appear in red.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
®
CMP 200
X-Ray Generator Service Manual
Figure 3C-35: Upgrade Media Selection window-'Choose a file' pop-up