Interfacing and Programming
3C.13.6 Touchscreen Setup (Cont)
Step
9.
Set the Exposure Volume by dragging the slider to the left (lower) or right (louder). This sets the loudness of the exposure active
tone.
10.
Set the LCD Brightness by dragging the slider left (dimmer) or right (brighter). This sets the brightness of the LCD display.
11.
Press OK to continue or CANCEL to return to the system utilities menu without making any changes. If OK was pressed, a pop-
up window will display asking if you wish to save the changes. Press YES to save the changes. Press NO to return to the system
utilities menu; CANCEL cancels the changes.
Features
Use these steps to select the digital functions.
Step
1.
From the touchscreen setup window, select FEATURES. Available digital interfaces will be shown on this screen.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-102
Figure 3C-31: Touchscreen setup window, Features tab
CMP 200
Action
Action
®
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C