CPI CMP 200 Series Service Manual page 117

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

3C
Interfacing and Programming
3C.7.3 Receptor Setup (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
AEC BACKUP
AEC BACKUP
MAS
AEC BACKUP MS
RAD HANG
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-48
FUNCTION
DESCRIPTION
(GenWare®)
AEC Mode
Defines the AEC backup mode to be used.
(AEC Back-up
FIXED:
Mode)
MAS:
MS:
Sets the maximum AEC backup mAs, to a limit of 600 mAs.
AEC Back-Up mAs
(AEC Backup mAs)
AEC Back-Up ms
Sets the maximum AEC back-up ms.
(AEC Backup ms)
Sets the time that the rotor will continue to spin after a rad exposure has terminated.
Rad Hangover (s):
The generator will determine the maximum AEC backup time, not to
exceed preset AEC backup mAs/ms values or system limits. The
characters AEC will be displayed in the time window of the LED
display during AEC operation.
Allows the operator to adjust the AEC backup mAs, not to exceed
preset AEC backup mAs/ms values or system limits. The mAs value
will be displayed in the time window of the LED display during AEC
operation.
Allows the operator to adjust the AEC backup ms, not to exceed preset
AEC backup mAs/ms values or system limits. The ms value will be
displayed in the time window of the LED display during AEC operation.
®
CMP 200
X-Ray Generator Service Manual
CPI Canada Inc.
Ch # 901476-04
Rev. C