CPI CMP 200 Series Service Manual page 249

X-ray generator
Hide thumbs Also See for CMP 200 Series:
Table of Contents

Advertisement

Reproducibility, Linearity, & HVL Testing
4.0
LINEARITY (Cont)
Record the mAs and K values taken from tables 3, 4, 6, and 7 below.
Table 3
Table 4
Table 6
Table 7
X
=
3
X
=
6
X
L =
X
In the numerator of the above equation, use the absolute value of X
(disregard the minus sign).
X
=
4
X
=
7
X
L =
X
In the numerator of the above equation, use the absolute value of X
(disregard the minus sign).
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 8
mAs
= _____
3
mAs
= _____
4
mAs
= _____
6
mAs
= _____
7
K
3
= ________
mAs
3
K
6
= ________
mAs
6
- X
3
6
= ________ (not to exceed 0.095)
+ X
3
6
Table 8: Linearity
K
4
= ________
mAs
4
K
7
= ________
mAs
7
- X
4
7
= ________ (not to exceed 0.095)
+ X
4
7
Table 9: Linearity
Rev. C
K
= _____
3
K
= _____
4
K
= _____
6
K
= _____
7
X-Ray Generator Service Manual Supplement
CPI Canada Inc
- X
3
6
- X
4
7
740917-00