6
Regular Maintenance
6.9.0
TUBE CONDITIONING / SEASONING (Cont)
2.
Season the tube at 70 kV by taking approximately 10 exposures of 200 mA and 100 ms. These
exposures should be taken at the rate of approximately one every 15 seconds.
3.
Restart the auto calibration sequence and manually terminate the exposure at 100 kV and 250 mA.
4.
Season the tube at 100 kV by taking approximately 5 exposures of 200 mA and 100 ms. These
exposures should be taken at the rate of approximately one every 15 seconds.
5.
Restart the auto calibration sequence and manually terminate the exposure at 120 kV and 160 mA.
6.
Season the tube at 120 kV by taking approximately 5 exposures of 160 mA and 100 ms. These
exposures should be taken at the rate of approximately one every 15 seconds.
7.
Restart the auto calibration sequence and allow the auto calibration sequence to complete.

6.10.0 END OF PRODUCT LIFE

The generator's useful life has been estimated at 10 years from point of sale. This will vary depending on use
and environmental conditions. If the generator has completed its useful service life, local environmental
regulations must be complied with in regard to disposal of possible hazardous materials used in the
construction of the generator.
In order to assist with this determination, the noteworthy materials used in the construction of this
generator are itemized below:
Electrical insulating oil in HT tank. This is a mineral oil with trace additives (25 Litre (6.5 U.S. gal)
Solder (lead/tin).
Epoxy fiberglass circuit board materials, tracks are solder on copper.
Wire, tinned copper. Insulated with PVC, tefzel, or silicone.
Steel and / or aluminum (generator cabinet and console chassis).
Plastic (console enclosure and console membrane).
Electrical and electronic components: IC's, transistors, diodes, resistors, capacitors, etc.
WARNING:
DO NOT DISASSEMBLE, INCINERATE, OR SHORT-CIRCUIT THE BATTERY(S) IN THIS
PRODUCT. DO NOT PUT IT IN TRASH THAT IS DISPOSED OF IN LANDFILLS; DISPOSE OF IT
AS REQUIRED BY LOCAL ORDINANCES.
THE FLUORESCENT LAMP IN THE LCD DISPLAY CONTAINS MERCURY. DO NOT PUT IT IN
TRASH THAT IS DISPOSED OF IN LANDFILLS; DISPOSE OF IT AS REQUIRED BY LOCAL
ORDINANCES.
THE LCD IS MADE OF GLASS. IF THE LCD BREAKS DUE TO ROUGH HANDLING OR
DROPPING, AND THE INTERNAL FLUID GETS IN YOUR EYES OR ON YOUR HANDS,
IMMEDIATELY WASH THE AFFECTED AREAS WITH WATER FOR AT LEAST 15 MINUTES.
SEEK MEDICAL ATTENTION IF ANY SYMPTOMS ARE PRESENT AFTER WASHING.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 6-12
®
CMP 200
X-Ray Generator Service Manual
ITEM
CPI Canada Inc.
Ch # 901476-08